Abstract
Aiming at the problems of various nonlinear characteristics caused by fiber twist in the manufacturing and packaging process of interventional catheter shape sensor, this paper proposes a fiber grating shape sensor configuration and its packaging method. The HPO-BPNN algorithm is employed to predict discrete curvature and the direction angles of curvature vectors, thus avoiding complex sensor parameter calibration steps. Additionally, the Frenet-Serret equation is integrated to achieve three-dimensional posture reconstruction of the catheter. Through the multi-curvature calibration experiment of the designed sensor, the average error at the end of the 2D curves is 21.15mm (3.91%), and for the 3D curves, it is 26.43mm (4.89%), thereby verifying the feasibility of the data-driven shape reconstruction algorithm.
| Original language | English |
|---|---|
| Title of host publication | 2024 IEEE INTERNATIONAL CONFERENCE ON CYBERNETICS AND INTELLIGENT SYSTEMS, CIS AND IEEE INTERNATIONAL CONFERENCE ON ROBOTICS, AUTOMATION AND MECHATRONICS, RAM, CIS-RAM 2024 |
| Publisher | Institute of Electrical and Electronics Engineers Inc. |
| Pages | 63-68 |
| Number of pages | 6 |
| ISBN (Electronic) | 979-8-3503-6419-4 |
| ISBN (Print) | 979-8-3503-6420-0 |
| DOIs | |
| State | Published - 8 Aug 2024 |
Publication series
| Name | IEEE Conference on Cybernetics and Intelligent Systems |
|---|---|
| ISSN (Print) | 2326-8123 |
Funding
Research supported by the Key R&D Program of Hubei Province under Grant 2022BAA066, and the National Natural Science Foundation of China under Grant 52275541.
Indexed by
- CPCI-S
- Scopus
Fingerprint
Dive into the research topics of '3D Shape Reconstruction Algorithm for Interventional Catheter Based on HPO-BPNN'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver