Abstract
A novel hybrid instrument of contact and non-contact measurement with large range is developed, and both measurement systems are based on a Linnik interference microscope and on white-light interference measuring techniques. The ambiguity presented in conventional monochromatic interferometers is not present in the contact and non-contact measurement, and they have a virtually unlimited unambiguous range. For the contact measurement, the vertical measuring range is ±5 mm with a resolution of 1 nm and the horizontal measuring range is ±25 mm in x-range and y-range with a resolution of 1.25 μm; for the non-contact measurement, the vertical measuring range is ±5 mm with a resolution of 1 nm and the horizontal resolution better than 0.5 μm. © 2007 Wuhan University.
| Original language | English |
|---|---|
| Pages (from-to) | 1066.0-1072.0 |
| Number of pages | 7 |
| Journal | Wuhan University Journal of Natural Sciences |
| Volume | 12 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jan 2007 |