Abstract
A device with tungsten-coil electrothermal vaporization for sample introduction into ICP has been proposed. It was applied to the determination of trace amounts of rare earth and other elements in rice samples. Several influencing factors were investigated in detail, such as drying and vaporization parameters, carrier gas flow rate, volume of vaporization chamber and matrix effects. Under optimal experimental conditions, the detection limits for Mg, Cu, Mn, Cr, Fe, Co, Ni and eight rare earth elements are of the order of 10-9-10-11 g. The detection limits for the rare earth elements tested by the present method are comparable to and, in most instances, exceed those for the GFAAS and conventional pneumatic nebulisation-ICP-AES. A precision with RSD<6% was obtained. © 1992 Springer-Verlag.
| Original language | English |
|---|---|
| Pages (from-to) | 54-58 |
| Number of pages | 5 |
| Journal | FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY |
| Volume | 344 |
| Issue number | 1-2 |
| DOIs | |
| State | Published - Jan 1992 |
Indexed by
- Scopus
Fingerprint
Dive into the research topics of 'Determination of trace amounts of rare earth and other elements in rice samples by ICP-AES with sample introduction by tungsten-coil electrothermal vaporization'. Together they form a unique fingerprint.Cite this