Abstract
A method for the determination of trace amounts of silver with a chemically modified carbon paste electrode is described. The modified electrode is prepared by simply mixing a chelating resin (a polythioether backbone and dioxymonosulphur polyethylene polyimines in the side-chain polymer) with graphite powder and Nujol oil. By immersing the electrode in a silver sample solution (pH = 6.5-7.5), silver can be adsorbed on the electrode surface and then determined by voltammetry in a separate blank solution. The response depends on the concentration of silver and the preconcentration time. For a preconcentration time of 5 min, the detection limit is about 3 × 10-10 M and the linear range is from 5 × 10-10 to 1 × 10-7 M with a relative standard deviation of 4%. Many common metal ions have no or little effect on the determination of silver. The recommended procedure was applied to the determination of trace amounts of silver in waste water. © 1990.
| Original language | English |
|---|---|
| Pages (from-to) | 213-219 |
| Number of pages | 7 |
| Journal | Analytica Chimica Acta |
| Volume | 229 |
| Issue number | C |
| DOIs | |
| State | Published - Jan 1990 |
Indexed by
- Scopus
Fingerprint
Dive into the research topics of 'Determination of trace amounts of silver with a chemically modified carbon paste electrode'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver