Abstract
Polytetrafluoroethylene (PTFE) was used as a fluorinating agent in order to facilitate a direct analysis of Cr in biological materials by electrothermal vaporization-inductively coupled plasma atomic emission spectrometry. The electrothermal vaporization parameters and the PTFE amount were optimized experimentally. The effects of different matrices on the emission signal of Cr were investigated in detail. Under the optimal conditions, the detection limit of Cr is 1.4 ng ml-1 and RSD 3.8% at 0.03 pg ml-1. The proposed method has advantages of sensitivity, simplicity, rapidity and no-chemical pretreatment, and has been applied to a direct determination of trace Cr in serum. © 1991, The Japan Society for Analytical Chemistry. All rights reserved.
| Original language | English |
|---|---|
| Pages (from-to) | 433-436 |
| Number of pages | 4 |
| Journal | Analytical Sciences |
| Volume | 7 |
| Issue number | 3 |
| DOIs | |
| State | Published - Jan 1991 |
Indexed by
- Scopus
Fingerprint
Dive into the research topics of 'Fluorination Assisted Electrothermal Vaporization-Inductively Coupled Plasma Atomic Emission Spectrometry for a Direct Determination of Chromium in Biological Materials'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver