Abstract
A pH ultramicrosensor based on polyaniline film modified carbon fiber microcylin-derelectrode in proposed. The film is introduced onto carbon fiber surface by electropolymerization of aniline. The sensor exhibits a linear response over the range 2∼12. 5 pH, with a near Nerstain response slope(60 mV/pH) at room Temperature. The correlation coefficient is 0. 999 9. The average response time at the sensor is about 2 min. The selectivity coefficient obtained for lgK H pot+, M n+ are -11.9, -12.1, -12. 0 and -11. 6 for K +, Na +, Li + and Ca 2+, respectively. The sensor was employed for measuring the pH of tap water, lake water and fresh urine. The results obtained from the sensor are in good agreement with those from the glass pH electrode. The sensor has been used for in vivo detection of pH on surface microenvironment of Brassica stigmata and pollen grains with satisfactory results.
| Translated title of the contribution | Investigations on carbon fiber pH ultramicrosensor modified by polyaniline film and its applications to the in vivo detection on brassica stigmata |
|---|---|
| Original language | Chinese |
| Pages (from-to) | x15-228 |
| Journal | Gaodeng Xuexiao Huaxue Xuebao/Chemical Journal of Chinese Universities |
| Volume | 18 |
| Issue number | 2 |
| State | Published - Jan 1997 |