Abstract
A desolavation device was connected to a microcolumn (40 mm × 3.0 mm i.d.) in which 8-hydroxyquinoline-5-sulphonic acid was immobilized on the substrate of active carbon-silica gel. The analyte and 1.0 M sodium acetate-acetic acid buffer solution (pH 5.6) were mixed and pumped through the microcolumn. The analyte was then preconcentrated and eluted from the column with 2.0 M hydrochloric acid. The efficiency of desolvation, improvement in sensitivity and signal stability over the temperature range 70-120°C were studied in detail. It was found that aluminium, copper, cadmium, iron and manganese could be separated from various co-existing elements with a sample volume of 1.2. ml and a carrier flow-rate of 1.2 ml min-. The desolvation efficiency at a desolvation temperature of 120°C 73%. The column could be used more than 100 times. The detection limits for the above metal ions were 0.34-5.1 ng ml-1 and the relative standard deviations (n=6) were 2.6-4.3%. The reliability of the method was confirmed for the analysis of human serum, and the results were in good agreement with those given by graphite furnace atomic absorption spectrometry. © 1993.
| Original language | English |
|---|---|
| Pages (from-to) | 887-893 |
| Number of pages | 7 |
| Journal | Analytica Chimica Acta |
| Volume | 283 |
| Issue number | 2 |
| DOIs | |
| State | Published - Jan 1993 |
Indexed by
- Scopus
Fingerprint
Dive into the research topics of 'On-line microcolumn preconcentration with desolvation and determination of trace elements by flow-injection inductively coupled plasma atomic emission spectrometry'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver