摘要
Small angle X-ray scattering (SAXS) is a widely used physical technique to characterize nanoscale materials. The absolute scattering intensity provides quantitative data to characterize the mass and density of the analyzed materials. However, the derivation of absolute scattering intensity usually requires the calibration of experimentally measured relative scattering intensity. The most common approach calibration employs standards with known differential cross sections. It is necessary to analyze the standard and samples under identical conditions, including the same distance from sample to detector. This study eliminates this limitation, i.e., the calibration may be achieved by measuring the standards and samples at different distances to the detector using variable sample locations.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 702-710 |
| 页数 | 9 |
| 期刊 | Instrumentation Science and Technology |
| 卷 | 52 |
| 期 | 6 |
| 早期在线日期 | 2月 2024 |
| DOI | |
| 出版状态 | 已出版 - 1 11月 2024 |
基金资助
No Statement Available
关键词
- Absolute intensity
- glassy carbon
- sample to detector distance
- small angle X-ray scattering (SAXS)
收录
- EI
- SCIE
- Scopus
学术指纹
探究 'Absolute intensity calibration of samples at variable sample to detector distances in small angle X-ray scattering (SAXS)' 的科研主题。它们共同构成独一无二的学术指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver