跳到主要导航 跳到搜索 跳到主要内容

Deep Learning-Based Model for Automatic Assessment of Anterior Angle Chamber in Ultrasound Biomicroscopy

  • Weiyan Jiang
  • , Yulin Yan
  • , Simin Cheng
  • , Shanshan Wan
  • , Linying Huang
  • , Hongmei Zheng
  • , Miao Tian
  • , Jian Zhu
  • , Yumiao Pan
  • , Jia Li
  • , Li Huang
  • , Lianlian Wu
  • , Yuelan Gao
  • , Jiewen Mao
  • , Yuyu Cong
  • , Yujin Wang
  • , Qian Deng
  • , Xiaoshuo Shi
  • , Zixian Yang
  • , Siqi Liu
  • Biqing Zheng, Yanning Yang*
*此作品的通讯作者

科研成果: 期刊稿件 › 论文 › 同行评审

17 引用 (Scopus)

摘要

Objective: The goal of the work described here was to develop and assess a deep learning-based model that could automatically segment anterior chamber angle (ACA) tissues; classify iris curvature (I-Curv), iris root insertion (IRI), and angle closure (AC); automatically locate scleral spur; and measure ACA parameters in ultrasound biomicroscopy (UBM) images.Methods: A total of 11,006 UBM images were obtained from 1538 patients with primary angle-closure glaucoma who were admitted to the Eye Center of Renmin Hospital of Wuhan University (Wuhan, China) to develop an imaging database. The UNet++ network was used to segment ACA tissues automatically. In addition, two support vector machine (SVM) algorithms were developed to classify I-Curv and AC, and a logistic regression (LR) algorithm was developed to classify IRI. Meanwhile, an algorithm was developed to automatically locate the scleral spur and measure ACA parameters. An external data set of 1,658 images from Huangshi Aier Eye Hospital was used to evaluate the performance of the model under different conditions. An additional 439 images were collected to compare the performance of the model with experts.Results: The model achieved accuracies of 95.2%, 88.9% and 85.6% in classification of AC, I-Curv and IRI, respectively. Compared with ophthalmologists, the model achieved an accuracy of 0.765 in classifying AC, I-Curv and IRI, indicating that its high accuracy was as high as that of the ophthalmologists (p > 0.05). The average relative errors (AREs) of ACA parameters were smaller than 15% in the internal data sets. Intraclass correlation coefficients (ICCs) of all the angle-related parameters were greater than 0.911. ICC values of all iris thickness parameters were greater than 0.884. The accurate measurement of ACA parameters partly depended on accurate localization of the scleral spur (p < 0.001).Conclusion: The model could effectively and accurately evaluate the ACA automatically based on fully automated analysis of UBM images, and it can potentially be a promising tool to assist ophthalmologists. The present study suggested that the deep learning model can be extensively applied to the evaluation of ACA and AC-related biometric risk factors, and it may broaden the application of UBM imaging in the clinical research of primary angle-closure glaucoma.
源语言英语
页(从-至)2497-2509
页数13
期刊Ultrasound in Medicine and Biology
卷49
期12
早期在线日期10月 2023
DOI
出版状态已出版 - 12月 2023

基金资助

The authors thank all of trainees and clinical collaborators for their contributions. This work was supported by the Major Research Project of Hubei Province (Grant No. 2020BCB055 to Y. Yang) and the National Natural Science Foundation of China (Grant No. 82101081 to S.W.) .

关键词

  • Deep learning
  • Ultrasound biomicroscopy
  • Ophthalmology
  • Primary angle-closure glaucoma
  • Anterior chamber angle

收录

  • Scopus
  • EI
  • SCIE

学术指纹

探究 'Deep Learning-Based Model for Automatic Assessment of Anterior Angle Chamber in Ultrasound Biomicroscopy' 的科研主题。它们共同构成独一无二的学术指纹。

引用此