By using a polytetrafluorethylene (PTFE) emulsion as fluorinating reagent, a method for fluorinating assisted eletrothermal vaporization (ETV)/ICP-AES direct determination of trace yttrium in TiO2 ceramic powder has been proposed; the various facts which would affect the vaporization process between the matrix and the analyte were studied; the fluorinating vaporization behavior of the matrix and the analyte both in slurry and in solution were investigated. Based on pre-separation between the matrix and the analyte, the matrix effect was reduced. On the optimum experiment conditions, the detection limit of the present method is 0. 26 mu g/L; and the RSD is 3. 8% (n = 5, c = 0. 5 mg/L). This method has been applied to directly determinate trace yttrium in two kinds of TiO2 powder without any chemical pretreatment.
| 投稿的翻译标题 | Direct determination of trace impurity yttrium in titantium oxide powder by fluorination assisted ETV-ICP-AES using slurry sampling |
|---|---|
| 源语言 | 中文 |
| 页(从-至) | 694-697 |
| 页数 | 4 |
| 期刊 | Gaodeng Xuexiao Huaxue Xuebao/Chemical Journal of Chinese Universities |
| 卷 | 21 |
| 期 | 5 |
| 出版状态 | 已出版 - 5月 2000 |
关键词
- fluorination electrothermal vaporization
- slurry sampling
- ICP-AES