A new method using a microcolumn packed with nanometer TiO2 as solid-phase extractant has been developed for the simultaneous preconcentration of trace amounts of Cu, Mn, Cr and Ni prior to their measurements by inductively coupled plasma atomic emission spectrometry (ICP-AES). Effects of pH, sample flow rate and volume, elution solution and interfering ions on the recovery of the analytes have been investigated. The adsorption capacity of nanometer TiO2 was found as 0.108, 0.149, 0.039 and 0.034 mmol g-1 for Cu, Cr, Mn and Ni, respectively. The separation of analytes can be achieved from water samples with a concentration factor of 50 times. The method has been applied for the determination of trace elements in biological sample and lake water with satisfactory results. © 2001 Elsevier Science B.V. All rights reserved.
源语言英语
页(从-至)207-213
页数7
期刊Analytica Chimica Acta
卷440
期2
DOI
出版状态已出版 - 1月 2001

基金资助

Wuhan University, WHU

关键词

  • chromium
  • copper
  • manganese
  • nickel
  • titanium dioxide
  • water
  • adsorption
  • article
  • atomic emission spectrometry
  • chemical analysis
  • flow rate
  • pH
  • priority journal
  • solid phase extraction

收录

  • Scopus

学术指纹

探究 'Nanometer-size titanium dioxide microcolumn on-line preconcentration of trace metals and their determination by inductively coupled plasma atomic emission spectrometry in water' 的科研主题。它们共同构成独一无二的学术指纹。

引用此