摘要
The processing and error analysis of signals in flow-injection systems were systematically studied by simulation and experimental measurements. The content includes an error analysis for peak-height and peak-area signal, a least-squares filtering procedure applied to the flow-injection curve and a peak recognition to remove interferences from air bubbles. Simulation results were obtained by statistical processing of peak-height and peak-area values from Gaussian curves to which noise had been added. The experimental measurements were done by an automatic flow-injection device to obtain detailed information for each individual point of a peak. 2-(2-Arsenophenylazo)-7-(2,6-dichlorophenylazo-4-sulphonic acid)-1,8-dihydroxynaphthalene-3,6-disulphonic acid (DCSA) was used for measuring physical dispersion alone, and Fe(II)-o-phenanthroline for the measurement of both physical dispersion and chemical reaction. The results from computer simulation and experiments agreed well. © 1990.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 337-342 |
| 页数 | 6 |
| 期刊 | Analytica Chimica Acta |
| 卷 | 235 |
| 期 | C |
| DOI | |
| 出版状态 | 已出版 - 1月 1990 |
基金资助
Chinese Universities Scientific Fund
关键词
- article
- flow injection analysis
- methodology
- nonhuman
- theoretical study
收录
- Scopus
学术指纹