跳到主要导航 跳到搜索 跳到主要内容

Processing and error analysis of signals in flow-injection analysis

  • Chen Danhua
  • , Zeng Yun'e

科研成果: 期刊稿件 › 论文 › 同行评审

16 引用 (Scopus)

摘要

The processing and error analysis of signals in flow-injection systems were systematically studied by simulation and experimental measurements. The content includes an error analysis for peak-height and peak-area signal, a least-squares filtering procedure applied to the flow-injection curve and a peak recognition to remove interferences from air bubbles. Simulation results were obtained by statistical processing of peak-height and peak-area values from Gaussian curves to which noise had been added. The experimental measurements were done by an automatic flow-injection device to obtain detailed information for each individual point of a peak. 2-(2-Arsenophenylazo)-7-(2,6-dichlorophenylazo-4-sulphonic acid)-1,8-dihydroxynaphthalene-3,6-disulphonic acid (DCSA) was used for measuring physical dispersion alone, and Fe(II)-o-phenanthroline for the measurement of both physical dispersion and chemical reaction. The results from computer simulation and experiments agreed well. © 1990.
源语言英语
页(从-至)337-342
页数6
期刊Analytica Chimica Acta
卷235
期C
DOI
出版状态已出版 - 1月 1990

基金资助

Chinese Universities Scientific Fund

关键词

  • article
  • flow injection analysis
  • methodology
  • nonhuman
  • theoretical study