摘要
Objectivet: To measure the performance of the two-dimension bone densit-ometer. Methods: BH 6012 Mode Bone Densitometer was used to do two-dimension scanning, and radioactivity through the bone was turned into the image by computer data processing. According to the structure and characteristics of the radius in the image, BMC and BMD in the "interest" area were present. Results:This instrument is accurate. Its bias is only 1.5% .It is precise. CV of BMC and BMD of the ultradistal radius of dry samples in a batch or between batches is 2.0% , in vivo 3.0% . Its linearity and stability are satisfactory. Conclusion : This Densitometer possesses excellent performance and provides a good means for the measurement of BMC and BMD of the ultradistal radius.